| Sample Conditioning |
| Filtration: | Heated unit with replaceable element 180 °C / 356 °F suitable for most applications 280 °C / 536°F for applications where ammonia is present in the flue gas |
| Filter Element: | 1.0 µm ceramic |
| Sample Cooler: | 2-stage cooler system (incorporating Peltier unit) with automatic condensate removal |
Blow-back
|
| Method: | Heated, filtered, compressed air |
| Triggers: | Internal timer; external trigger (via digital link); Manual (push button); +24 V external drive |
| Temperature: | 150 °C / 300 °F |
| Timer Interval: | 15 minutes to 24 hours user selectable |
| Duration: | 3 secs |
| Control System |
| Communications: | Remote control via RS 485 digital link |
| Automatic Controls: | Filter temperature; Cooler temperature; Condensate removal; Blow-back timing; Blow-back air temperature |
| Manual Controls: | Blow-back trigger |
| Status LEDs: | Power; Blow-back; System OK; Communications |
Status Outputs (via digital link): | Filter temperature; Blow-back air temperature; Blow-back pressure; Cooler Temperature; Enclosure temperature; Condensate pumps active; Sample gas temperature (optional) |
| Alarms (indicated at analyzer): | Filter temperature (low/high) Blow-back accumulator temperature (low/high) Cooler temperature (low/high) Blow-back accumulator low pressure Moisture slip Pump fault |
| Relay Outputs: | System OK; Blow-back in progress; Filter cold; Cooler fault |
| Additional Output: | Power for anti-freeze condensate sample line |
| Environmental | |
| Enclosure: | Stainless steel sealed to the requirements of IP 65 / NEMA 4X |
| Operating Temperature: | -25 to + 55 °C / -13 to +131 °F (option to -40 °C / -40 °F) |
| Compliance |
| EMC: | Conforms to EN-61326 (Class A) |
| Electrical Safety: | Conforms to EN-61010-2 |
| Power |
| Power Supply: | 110 or 230 V a.c., ±10%, 50 or 60 Hz |
| Power Consumption: | 900 W |
| Air Requirements |
| Air Requirements for cooling: | Dependent upon ambient temperature and process conditions |
Dimensions (H x W x D) | 685 x 510 x 356 mm / 27 x 20 x 14 in |
| Weight: | 56 kg / 124 lb |
| Mounting: | Single flange - 3" ANSI, 150 lb type |
| Flange Temperature: | up to 180 °C / 356 °F without adaptor up to 500 °C / 932 °F with adaptor |
Optional Flange Adaptor: | 2", 3", or 4" ANSI 150 lb; DN65 PN6 |
| Probe Pipes and Materials |
| 316 Stainless Steel : | up to 600 °C / 1112 °F |
| Hastelloy C-276: | up to 900 °C / 1652 °F |
| Hastelloy X: | up to 1200 °C / 2200 °F |
| Mullite: | up to 1700 °C / 3100 °F |
| Lengths*: | 0.5 m / 1.6 ft; 1.0 m / 3.3 ft; 1.5 m / 5 ft; 2.0 m / 6.6 ft |
| Probe Pre-filter#: | 2.0 µm Sintered stainless steel (Temp. <400 °C / 750 °F) 2.0 µm Hastelloy (Temp. up to 800 °C / 1470 °F) |
* Mullite probes only available in 0.5 and 1.0m lengths # Available with all probe types except mullite: |
| Options |
| Sample Gas Cooler Temp. sensor: | Type K thermocouple |
| Air blower for instrument cooling: | Low capacity or high capacity blower units |
| Continuous Product Development may make it necessary to change these details without notice |
| |